Features:
1.The microcomputer intelient.ull-automatic controlis adopted,and the vhole process of heating, sterilization,exhaust and drying is automatcally con-trolled and swiched,.Users can set their own programs for sterilization temperature, stermlization time and drying time according to the needs of steiliedarticles, without supervision in the whole process.
2.Full-automatic water adding, water shortage automatic water replenishing function.Cut of the heating power supply due to water shortage, and auto-matically resume working anter water replenishment ithas the functions of preventing dry burning and over-temperature automatic protection. When thewater level is lower than the low water level or exceeds the set temperature,the system automatcally cuts of the power supply and gives an auible andvisual alarm.
3.(Pre-selected)Being equipped with buli-in micro thermal printer to realize data traceability and real-time priting of operating parameters.
4. in the inner chamber with double-layer structure,the door pane is made of SUs304 matera, which is polished mechanically, smooth ft,free of deadangle and easy to clean.
5.The new structure realizes the front and rear doors (pre-selected) and the sealed external stucture, which can separate the bactera areafrom thesterile area,and better ensure that the sterilized instruments are not polluted.
6.The door is equipped with a safety interock devic.When there is pressure in the steiliation chamber , the door is automatically locked and cannot beopened to ensure the safety of personel.There is also a door control swich. Only when the doo is closed in place can the power be turned on.
7.Equioped with overpressure protection device: when the pressure exceeds the working pressure,the safety vave ilimmediately release the pressurefor protection.
8,Over-current protection device, when the current overload, the power circuit breaker will automatically cut of the power supply inmediately.9.The sterilizer is with the jacketed design, which has the drying function and achieves the drying function of sterilized articles.
10.Sterilization modes can be divided into three types:(1) Sterilization at 121ºC;(2) Sterlization at 132ºC;(3) Self-set.
*The main body is in a hoizontal and double-layer stucture,the imer chamber uses stainless steel materialsof high qualty it s resistant to corrosion and has adurable quality. The fabrication conforms with the quality and safety standard of the class l pressure vessels.
*The air-tightness adopts the handwhee rotaig structre, an it aso equigs the salet interock of uick-opening dox,these features make t o be sale and reliabe.
*it owns double overpressure protection function,the precise pressure controller and automatic overpressure steam-release vave ill uarantee the salety ofpersonnel, equipment and materials.
*This equipment has drying function, automatically control the warning function of water shortage.
*The pressure can be adjusted from 0.10Mpa to 0.22Mpa, and the temperature is adjustable from 115ºC to 135ºC.The air-tight door has two types: single gate or double gate.
*Al he major parts are purchased from famous brands at home and broad, which ensure the safety reiailiy and maneuwverability of the sterilization equipment.
Model | HE-F300 | HE-F430 | HE-F600 | HE-F800 | HE-F1000 | HE-F1200 | HE-F1500 |
Design pressure (Mpa) | 0.25 | 0.25 | 0.25 | 0.25 | 0.25 | 0.25 | 0.25 |
Design temperature | 138 | 138 | 138 | 138 | 138 | 138 | 138 |
Working pressure (ºC) | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 | 0.22 |
Working temperature(ºC) | 135 | 135 | 135 | 135 | 135 | 135 | 135 |
Temperature adjustment range (ºC) | ±2 | ±2 | ±2 | ±2 | ±2 | ±2 | ±2 |
Air pressure | 0.3-0.6Mpa | 0.3-0.6Mpa | 0.3-0.6Mpa | ||||
Power supply | Three-phase AC380V | Three-phase AC380V | Three-phase AC380V | Three-phase AC380V | AC220V | AC220V | AC220V |
Power | 18 | 24 | 30 | 40 | 1 | 1 | 1 |
Sterilization time setting range (minutes) | 0~60 | 0~60 | 0~60 | 0~60 | 0~60 | 0~60 | 0~60 |
Sterilization working temperature(ºC) | 115-135ºC adjusting | 115-135ºC adjusting | 115-135ºC adjusting | 115-135ºC adjusting | 115-135ºC adjusting | 115-135ºC adjusting | 115-135ºC adjusting |
Setting scope of drying time(minutes) | 0~60 | 0~60 | 0~60 | 0~60 | 0~60 | 0~60 | 0~60 |
Drying working temperature (ºC) | 80-100 | 80-100 | 80-100 | 80-100 | 80-100 | 80-100 | 80-100 |
Capacity | 0.3 m³(300L) | 0.43 m³(430L) | 0.6m³(600L) | 0.8m³(800L) | 1m³(1000L) | 1.2m³(1200L) | 1.5m³(1500L) |
Dimension of sterilization chamber(mm) | 900*550*650 | 1000 *630 *750 | 1350*630*750 | 1570*650*800 | 1200 *900 *1000 | 1400*900 *1000 | 1700 *900 *1000 |
Overall dimension (mm) full-automation | 1270*820*1700 | 1450 *860 *1850 | 1920*860 *1850 | 2180 *970 *1950 | 1800 *1300 *1950 | 2000 *1300*1950 | 2250 *1220 *1900 |
Weight(full-automation) | 560 | 630 | 800 | 1000 | 1100 | 1200 | 1300 |